Accuracy and reliability of a wireless vital signs monitor for hospitalized patients in a low-resource setting

dc.contributor.authorNantume, Assumpta
dc.contributor.authorKiwanuka, Noah
dc.contributor.authorMuyinda, Asad
dc.contributor.authorCauvel, Teresa
dc.contributor.authorShah, Sona
dc.date.accessioned2023-04-28T11:38:59Z
dc.date.available2023-04-28T11:38:59Z
dc.date.issued2022
dc.description.abstractThe purpose of this study was to evaluate the accuracy and reliability of neoGuard in comparison to a conventional bedside monitor on patients in a low-resource clinical setting. This was a single-arm methods comparison study involving the use of a wearable vital signs monitor (neoGuardTM) versus a conventional bedside monitor (Edan iM8). The study was conducted at Jinja Regional Referral Hospital, a tertiary care hospital situated in Eastern Uganda. Participants: Thirty patients (10 male, 20 female) were enrolled from the adult recovery ward at JRRH. Participants were eligible for the study if they were at least 18 years of age, had 2 sets of normal vital sign measurements obtained 1 h apart, and were able and willing to provide informed consent. Main Outcome and Measures: The primary outcome measures were (i) bias (mean deviation) and (ii) limits of agreement [95% CI]. Bland-Altman plots were generated to illustrate the level of agreement between the neoGuardTM technology and the Edan iM8 monitor. Bland-Altman analysis was performed for 24 participants; datasets from six participants were excluded due to missing or invalid measurements. Findings showed a moderate level of agreement for measurement of SpO2, PR, and RR, with >80% of subject means falling within the predefined acceptability limits. However, there was also notable variation in accuracy between subjects, with large standard deviations observed for measurement of all four parameters. While the level of agreement for measurement of temperature was low, this is partly explained by limitations in the comparison method.en_US
dc.identifier.citationNantume, A., Kiwanuka, N., Muyinda, A., Cauvel, T., & Shah, S. (2022). Accuracy and reliability of a wireless vital signs monitor for hospitalized patients in a low-resource setting. Digital Health, 8, 20552076221102262.DOI: 10.1177/20552076221102262en_US
dc.identifier.urihttps://nru.uncst.go.ug/handle/123456789/8573
dc.language.isoenen_US
dc.publisherDigital Healthen_US
dc.subjectdigital healthen_US
dc.subjectvital signsen_US
dc.subjectwearableen_US
dc.subjectnewborn healthen_US
dc.titleAccuracy and reliability of a wireless vital signs monitor for hospitalized patients in a low-resource settingen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Accuracy and reliability of a wireless.pdf
Size:
1.68 MB
Format:
Adobe Portable Document Format
Description:
Accuracy and reliability of a wireless vital signs monitor for hospitalized patients in a low-resource setting
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: